Dynascan LeakVIEW

LeakVIEW packaging seal integrity testing

LeakVIEW provides visual detection of leaks through either bubble emission or bust testing in a wide range of packaging. An essential piece of quality assurance equipment for testing the seal integrity for all types of packaging within the food, consumer and medical device industries.

The clear perspex test chamber is filled with water and during testing, the head-space is placed under a predetermined vacuum level causing the packing under test to inflate. Should any leaks be present a stream of bubbles will be emitted showing the location of the leak in a faulty seal. The system is able to identify holes as small as 20 microns (µm) and operates in compliance with ASTM D3078 or ASTM F2096 standards. For altitude simulation or burst testing water is not required in the test chamber.

The system requires no calibration or electrical supply and is easy to use by operators with minimum training providing dependable repeatable results.

LeakVIEW Automatic

An automatic version is also available that uses a programmable touch screen so that predetermined vacuum level and test duration can be undertaken at the touch of a button. This ensures total repeatability and reproducibility of results regardless of the operator skill. This version is used for altitude simulation testing. The display will indicate the end of the test cycle and in the case of altitude simulation, ascent and descent profiles can be displayed graphically.

Both versions are available in standard ranges of sizes to suit the application. In addition, custom configurations can be provided to suit special customer application.

Features

  • Standard and custom size chambers available
  • Bubble emission and altitude simulation testing
  • Low maintenance
  • Automated or manual control
  • Detect holes down to 20µm in size

Specification

LeakVIEW Bubble emission leak detection system Custom configurations available for chamber size and internal pressurisation
Dimensions A range of sizes are available, both standard and custom
Please contact us for further information
Limit of Detection Better than 20 microns hole diameter
Performance is a function of pack size, vacuum level and head space
Evacuation Time Example (Small Chamber) 0 to -800 mbar <30 seconds at 6 bar line pressure (empty)
0 to -200 mbar <5 seconds at 6 bar line pressure (empty)
Control LeakVIEW Manual Combination evacuation hold and vent control
LeakVIEW Automatic Touch screen and start-stop reset button
Indicators 0-11 bar pressure, 0-1 bar vacuum
Materials Acrylic test chamber Grade 304 stainless steel base
Compliance LeakVIEW Manual Can operate in compliance with ASTM D3078 or ASTM F2096
LeakVIEW Automatic Can operate in compliance with ASTM D3078 or ASTM F2096
Standards Directive 97/23/EC CAT S.E.P. applies
CE marked
Calibration No calibration required
Gauges may be verified for accuracy

Download LeakView Manual Datasheet
Download LeakView Automatic Datasheet